GNDE Products · Inspection Electronics
Pulser and Receiver Instruments
Magnetostrictive guided-wave inspection electronics. Explore transmitter, receiver, and multiplexer specifications alongside the corresponding instrument photographs.
MsSRv5 — Fifth-Generation Pulser/Receiver
Dual-channel instrument for axially symmetric magnetostrictive transduction probes; no 16-channel multiplexer.
Transmitter
- Two synchronous, differentially driven high-voltage burst outputs
- Square-wave excitation; 1–14 selectable cycles
- 1–500 kHz operating frequency
- Maximum output voltage: 240 Vpp; output current: 40 A peak-to-peak
- Selectable output amplitude: 0, 25, 50, or 100%
- Pulse repetition rate: 1, 2, 4, 8, or 16 pulses/s
Receiver
- Two high-voltage differential inputs; electronic ±90° phase shifting for directionality control
- Programmable gain: 0–80 dB in 1-dB steps; time-controlled gain: 0–80 dB linear ramp
- Eight 4-pole active filters: 16, 32, 45, 60, 90, 128, 180, and 250 kHz; other choices available
- Pulse-echo and pitch-catch modes

MsSRv5M — Sixth-Generation Pulser/Receiver
Instrument with 16-channel relay-type multiplexer for segmented magnetostrictive sensors and arrays.
Transmitter
- Two synchronous, differentially driven high-voltage burst outputs; in-phase or ±90° phase displacement
- Square-wave excitation; 1–14 selectable cycles
- Frequency: 1–500 kHz
- Maximum output: 240 Vpp and 40 A peak-to-peak
- Output amplitude: 0, 25, 50, or 100%
- Pulse repetition rate: 1, 2, 4, 8, or 16 pulses/s
Receiver
- Two differential inputs with phase adjustment for directionality
- Gain: 0–80 dB in 1-dB steps; time-controlled gain: 0–80 dB
- Eight selectable 4-pole active filters: 16, 32, 45, 60, 90, 128, 180, and 250 kHz
- Pulse-echo and pitch-catch modes
Multiplexer
16-channel relay-type multiplexer, configurable for transmitting and receiving across the 16 sensor channels.


Both instruments are designed and manufactured by Southwest Research Institute (SwRI).